User's Manual

USER MANUAL P/N UM1400-00-09 8/10/20 Pg 33
Recovery Force 10022 Lantern Rd., Suite 100, Fishers, IN 46037
E. Troubleshooting
NOTE: Before continuing, check visually for any defects.
Error Screen
Error Description
Corrective Action
Battery was removed during
operation.
Insert charged battery into the MAC System
Controller
An incorrect date has been set in
the Controller.
Press OK and then set correct date on
Controller
Date and/or time is not correct
on Controller.
Set correct date and time on Controller
A critical error has occurred with
the Controller.
Error Code:
000: No Error
001: Battery Low
002: Battery Critically Low
003: Battery Empty
004: Motor Running Error
005: I2C Communication Error
006: SPI Communication Error
007: UART Communication Error
008: ADC Sampling Error
009: PWM Output Error
010: Battery Unhealthy
011: Battery Over-Temperature
Power down the Controller. Replace the
Controller if the issue is not resolved by
powering down the device and rebooting.