D.3 Electromagnetic Interference
EMC Manufacturer Declarations
Movement and Compressions System - electromagnetic emissions - manufacturer declaration
The Movement and Compressions System is intended for use in the electromagnetic environment specified below. The
customer or the user of the Movement and Compressions System should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Movement and Compressions System uses RF energy only for its
internal function. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class BF
The Movement and Compressions System is suitable for use in all
establishments, including domestic establishments and those directly connected
to the public low voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage
fluctuations/flicker
emissions IEC 61000-3-3
Not applicable
Movement and Compressions System - electromagnetic immunity - manufacturer declaration
The Movement and Compressions System is intended for use in the electromagnetic environment specified below. The customer
or the user of the Movement and Compressions System should assure that it is used in such an environment. For Charging Hub.
Immunity Test
IEC 60601 test
level
Compliance level
Electromagnetic environment - guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
Complies
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30%.
Electrical fast
transient/burst IEC
61000-4-4
±2 kV for power
supply lines
Complies
Mains power quality should be that of a typical commercial,
home use or hospital environment.
Surge
IEC 61000-4-5
±1 kV line to line
±2 kV line to earth
Complies
Mains power quality should be that of a typical commercial,
home use or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
>95% dip in U
T
for
10ms
60% dip in U
T
for
100ms
30% dip in U
T
for
500ms
>95% dip in U
T
for
5000ms
Complies
Mains power quality should be that of a typical commercial,
home or hospital environment.