Datasheet
www.recom-power.com
REV.: 11/2019 MED-5
DC/DC Converter
Specifications (measured @ Ta= 25°C, nominal input voltage, full load and after warm-up)
REM6
Series
DIMENSION and PHYSICAL CHARACTERISTICS
Parameter Type Value
Material
case
PCB
potting
non-conductive black plastic (UL94-V2)
FR4 (UL94-V1)
silicone (UL94-V0)
Dimension (LxWxH) 31.80 x 20.30 x 10.40mm
Weight 14g
continued on next page
Notes:
Note6: An external input filter capacitor is required if the model has to meet EN61000-4-4 or/and EN61000-4-5
Recommended components: 5Vin aluminium capacitor (Nippon Chemi-con KY series, 1000µF/25V) and a
reverse diode (Vishay V10P45) to connect in parallel
12Vin, 24Vin aluminium capacitor (Nippon Chemi-con KY series, 470µF/50V)
48Vin aluminium capacitor (Nippon Chemi-con KY series, 330µF/100V)
Note7: The whole REM6 series can meet EMI Class A with no external filter. And Class B only with external components
MODEL
C1
(8)
C2
(8)
L1
(8)
REM6-05xxS_D 22µF/16V MLCC 22µF/16V MLCC 137µH
CMC
REM6-12xxS_D
REM6-24xxS_D(/W)
4.7µF/50V MLCC 4.7µF/50V MLCC 227µH
CMC
REM6-48xxS_D(/W) 2.2µF/100V MLCC 1µF/100V MLCC 419µH CMC
Certificate Type (Others) Conditions Standard / Criterion
Medical electrical equipment - Part 1-2: General requirements for basic safety and essential
performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
EN60601-1-2:2015
Industrial, scientific and medical equipment - Radio frequency disturbance character-
istics - Limits and methods of measurement
(7)
EN55011:2009 + A1:2010 Class A & B
ESD Electrostatic discharge immunity test Air ±15kV; Contact ±8kV EN61000-4-2:2008
Radiated, radio-frequency, electromagnetic field immunity test
10V/m (80-2500MHz)
27V/m (385MHz)
28V/m (450MHz)
EN61000-4-3:2006 + A2:2010
Fast Transient and Burst Immunity
(6)
DC Port: ±2kV EN61000-4-4:2012
Surge Immunity
(6)
DC Port: ±2kV EN61000-4-5:2005
Immunity to conducted disturbances, induced by radio-frequency fields
6Vr.m.s EN61000-4-6:2013
Power Frequency Magnetic Field
30A/m EN61000-4-8:2009
EMC Filter Suggestion for Class B
(8)
C1 C2
+V
in
-V
in
+V
out
-V
out
+V
in
-V
in
+V
out
-V
out
L1
Notes:
Note8: The component values can be adapted according to customer´s application