Datasheet

CTRL Option 
(4)
Pinning 
(3)
Input Type 
(2)
Model Numbering
Refer to Applications Notes
DC/DC Converter
Output Power
Input Voltage 
(1)
Output Voltage 
Single or Dual 
MED-1
REV.: 0/2015www.recom-international.com
Notes:
  Note1:  for 4:1 Input Voltage Type add “W”, see Note 2. 
       2:1   4:1 “W”
      xx= 4.5-9 Vin  =  “05”  xx= 9-36Vin  =  “24”
      xx= 9-18Vin  =  “12”  xx= 18-75Vin  =  “48”
      xx= 18-36Vin  =  “24”
      xx= 36-75Vin  =  “48”
 Note2:  Blank for Standard 2:1 Input Voltage Range; „W“ suffix for 4:1 Input Voltage Range
  Note3:  „A“ suffix for A pinning; „C“ suffix for C pinning, for more details refer to Package Style and Pinning
  Note4:  „CTRL“ suffix for control pin option, for A pinning only, for C pinning not aviable 
Examples:
REM10-0512D/A  =  2:1 Input,  4.5-9Vin,  ±12Vout,  pinout „A“,  without control pin
REM10-1215S/C  =  2:1 Input,  9-18Vin,  15Vout,  pinout „C“,  without control pin
REM10-4815SW/A/CTRL  =  4:1 Input,  36-75Vin,  15Vout,  pinout „A“  with control pin
REM10-243.3SW/C  =  4:1 Input,  9-36Vin,  3.3Vout,  pinout „C“,  without control pin
10 Watt 
2:1 & 4:1
DIP24
Single and Dual 
Output
3
Y
E
A
R
w
a
r
r
a
n
t
y
3
REM10- __ __ __ __ / __ __
Features
Regulated
Converters
• Reinforced Insulation for 250VAC Working Voltage
• Clearance and Creepage Distance: 8mm
• 5kVAC I/P to O/P 2MOPP Isolation
• 2µA Patient Leakage Current
• Industry Standard Pinout
• 2:1 and 4:1 Wide Input Range
Description
The REM10 series of medical grade regulated DC/DC converters feature reinforced 5kVAC/1 minute 
isolation with low 2µA leakage and are 60601-1 3rd Ed. certified for 250VAC continuous working. The 
compact DIP24 package offers tightly regulated single and dual outputs, even under no-load conditions. 
The outputs are short circuit and overload protected. The converters are available in two different pin-
ning options and optionally with an external control pin for standby consumption as low as 12.5mW. The 
converters are fully certified to CB, IEC/EN and ANSI/AAMI standards and carry the CE and UL marks.
Selection Guide
Part  Input  Output  Output  Efficiency  Max. Capacitive
Number  Voltage Range  Voltage  Current  typ.  Load
  (VDC)  (VDC) (mA)  (%)  (µF)
REM10-xx3.3S/*  4.5-9 / 9-18 / 18-36 / 36-75  3.3  2500  80 / 83 / 83 / 82.5  3000 
REM10-xx05S/*  4.5-9 / 9-18 / 18-36 / 36-75  5  2000  84 / 85.5 / 86.5 / 86.5  2500 
REM10-xx12S/*  4.5-9 / 9-18 / 18-36 / 36-75  12  830  86.5 / 88 / 89 / 89  430 
REM10-xx15S/*  4.5-9 / 9-18 / 18-36 / 36-75  15  670  87 / 89 / 89 / 89  350
REM10-xx24S/*  4.5-9 / 9-18 / 18-36 / 36-75  24  416  85.5 / 89 / 89 / 88.  125
REM10-xxx05D/*  4.5-9 / 9-18 / 18-36 / 36-75  ±5  ±1000  83 / 84 / 85 / 85  ±1440
REM10-xx12D/*  4.5-9 / 9-18 / 18-36 / 36-75  ±12  ±416  85.5 / 89 / 89 / 88  ±250 
REM10-xx15D/*  4.5-9 / 9-18 / 18-36 / 36-75  ±15  ±333  86.5 / 88 / 89 / 88  ±180 
REM10-xx3.3SW/*  9-36 / 18-75  3.3  2500  83 / 82.5  3000
REM10-xx05SW/*  9-36 / 18-75  5  2000  86.5 / 86.5  2500
REM10-xx12SW/*  9-36 / 18-75  12  830  89 / 89  430
REM10-xx15SW/*  9-36 / 18-75  15  670  89 / 89  350
REM10-xx24SW/*  9-36 / 18-75  24  416  89 / 88.5  125
REM10-xx05DW/*  9-36 / 18-75  ±5  ±1000  85 / 85  ±1440
REM10-xx12DW/*  9-36 / 18-75  ±12  ±416  89 / 88  ±250
REM10-xx15DW/*  9-36 / 18-75  ±15  ±333  88 / 88  ±180
REM10
2MOPP
250VAC
E224736
UL-60950-1 (Pending)
EN-55011 (Pending)
EN-55022 (Pending)
IEC/EN-60950-1 (Pending)
IEC/EN-60601-1 (Pending)