Datasheet

www.recom-power.com
REV.: 5/2018 MED-4
DC/DC Converter
Specifi cations (measured @ Ta= 25°C, nominal input voltage, full load and after warm-up)
REM1
Series
ENVIRONMENTAL
Parameter Condition
Value
Operating Temperature Range without derating (see graph) -40°C °C to +85°C
Maximum Case Temperature 105°C
Temperature Coeffi cient 0.02%/K typ.
Operating Altitude
according to IEC/EN60601-1
accroding to IEC/EN62368-1
3000m
5000m
Operating Humidity non-condensing 5% - 95% RH max.
Pollution Degree PD2
MTBF according to MIL-HDBK-217F, G.B.
+25°C
+85°C
18200 x 10
3
hours
7500 x 10
3
hours
Vibration according to MIL-STD-202G standard
100 12080
85 90
6040200-20-40
0
20
30
40
60
80
100
Ambient Temperature [°C]
Output Power [%]
Derating Graph
(@ chamber and natural convection 0.1m/s)
SAFETY AND CERTIFICATIONS
Certifi cate Type (Safety) Report / File Number Standard
Audio/video, information and communication technology equipment. Safety
requirements (CB Scheme)
L0339m31-A-L EN62368-1:2014
Medical Electric Equipment, General Requirements for Saftey and Essential
Performance E314885-D1000-1/A0/C0-UL
ANSI/AAMI ES60601-1:2005/
®
2012 +
A1:2012 + C1:2009/
®
2012 + A2:2010/
®
2012
CSA C22.2 No. 60601-1:14, 3rd Edition, 2014
Medical Electric Equipment, General Requirements for Safety and Essential
Performance
E314885-D1000-1/A0/C0-CB
IEC60601-1:2005 +AM1:2012
EN60601-1:2006 + A12:2014
EAC RU-AT.49.09571 TP TC 004/2011 TP TC 004/2011
RoHS2+ RoHS-2011/65/EU + AM-2015/863
EMC Compliance Condition Standard / Criterion
Medical electrical equipment - Part 1-2: General requirements for basic safety
and essential performance - Collateral standard: Electromagnetic compatibility
IEC60601-1-2:2014
EN60601-1-2:2015
Industrial, scientifi c and medical equipment - Radio frequency disturbance
characteristics - Limits and methods of measurement
with external fi lter EN55011, 2009+A1:2010, Class B
ESD Electrostatic discharge immunity test Air: ±15kV; Contact: ±8kV IEC61000-4-2:2008 , Criteria A
Radiated, radio-frequency, electromagnetic fi eld immunity test
3V/m IEC61000-4-3:2006+A2:2010, Criteria A
Fast Transient and Burst Immunity
DC Port: ±2kV IEC61000-4-4:2012, Criteria A
Surge Immunity
DC Port: ±1kV IEC61000-4-5:2014, Criteria B
Immunity to conducted disturbances, induced by radio-frequency fi elds
DC Port: 6V IEC61000-4-6:2013, Criteria A
Power Magnetic Field Immunity
50Hz, 30A/m IEC61000-4-8:2009, Criteria A
continued on next page