User Guide
COPD QUICK REFERENCE GUIDE
2
Example Preset 2:
High Flow Therapy
Setting Value
Mode SIMV-Pressure
Patient Type Adult
High Flow ON
Flow
2
30 L/min
Alarms
(set Alarms according to your
Protocol)
High Insp. Pres-
sure
Patient Circuit
Discon.
High Pressure
Delay
Setting Value
Mode Vol. Targeted-PS
Patient Type Adult
Humidication HME or Humidier
Circuit Type Valveless
Breath Rate 12 BPM
Inspiratory
Time
1.0 Seconds
Tidal Volume 7-8 mL/kg (IBW)
Pres. Adj. Rate Slow
Pres. Minimum 8 cmH2O (above set PEEP)
PEEP 5 cmH2O
Flow Trigger 3.0 L/min
Flow Cycle
4
65%
Time Cycle 1.5 Seconds
Rise Time 2 or 3
Apnea Rate 10 Backup BPM
Alarms
(set Alarms according to your
Protocol)
Inspiratory
Pressure
(High) 35 cmH2O
1 - The settings referenced in this guide are suggestions and do not replace physician orders; please consult with patient
physician to make sure that all settings are appropriate for the patient.
2 - Adjust ow to achieve target SpO2 and/or CO2 targets.
3 - Adjust to patient comfort.
4 - See Flow Cycle Guide (pg. 4).
Example Preset 1:
Non-Invasive Ventilation
1, 3