User manual

ProgrammingGuide
74 MI
XF factorydefaultvalueswillbeset
11.6.5LockMode
The access to the unit is possible via serial interface (software) and
viathedirectuserinput(modebuttons,LCDdisplay).Itispossible
tolockthebuttons.Thisallowsaccesstheunitonlyviasoftware.
J=L directuserinputvia
modebuttonsdenied
Alternatively the unit can be unlockedby pressing the <Mode/Up>
buttonssimultaneouslyforthreeseconds.
11.6.6ModeSettingfortheDigitalInputFTC3
Thedigital inputFTC3(see section
5.4.4TriggerandHoldFunction
onpage28)canbeusedasfollows:
XN=T FTC3astrigger
XN=H FTC3withholdfunction
11.6.7ChangingtheSensingHeadCalibrationData
If it is necessary to exchange the sensing head, you must set the
calibrationdataforthenewsensinghead:
XZ=01234567FFFFFFFF accordingtothe
headcalibrationdata
ForMID/MICmodelsthelasteightnumbersare notusedandmust
besetto„F“likeshownintheexampleabove.ForMIHmodelsand
specials(likeMTBorG5)allsixteennumbersareused.
11.6.8AmbientBackgroundTemperatureCompensation
In case of compensating the ambient
background temperature, the
followingmodesareavailable:
AC=0 nocompensation
AC=1 compensation with a constant temperature value set
withcommandA.