User manual

Installation
22 MI
5.3.3ThermocoupleOutput
If you are using a J‐ or K‐ thermocouple you must install a
compensation cable. The cable is available as an accessory
(XXXCI1CB25 for Type J, XXXCI2CB25 for Type K) with a cable
lengthof7.5m(24.6ft.)
Connectthewiresaccordingtothefollowingtable:
J-Thermocouple Power Supply
+
white
red-white
+
red-yellow
yellow
Table1:WiringtheThermocoupleJCompensationCable
K-Thermocouple Power Supply
+
yellow
red-yellow
+
red-white
white
Table2:WiringtheThermocoupleKCompensationCable
Figure12:WiringtheThermocoupleJCompensa tionCable
Power +
red-
y
ellow
Power
y
ellow
TC J
red-white
TC J +
white