Maintenance
52 MarathonSeriesFA/FR
Figure39:AttachingtheFiberOpticCabletotheElectronicsHousing
10.4.3FiberCalibration
Each replacement fiber optic cable is calibrated at the factory before shipping. The calibration
constants are sent along with a label mounted on the cable. So you have to enter them into the
appropriate Fiber Calibrationsoftware program. This program sends the newcalibration constants,
throughtheRS485connection,tothe
sensor’selectronics.
The Fiber Calibration program comes with the other software programs you received. To run the
programandenternewcablecalibrationconstants,completethefollowing:
1. The program can not be launched from the CD. Thus you have to copy the file
MARATHFC.EXEfromthesoftwareCDto
theharddiskofyourcomputer,e.g.bymeansof
theWindowsExplorer.
2. Forlaunchingtheprogramyouhavetoselectthefileandtopushthe<Enter>button.
3. InthefollowingdialogyouarerequestedtoselecttherightCOMportwiththepluggedunit.
Forestablishingthe
communicationclickonthe<Done>button.
4. Themainscreenappears.Clickonthe<FiberID>button.
5. Inthefollowingdialogyouarerequestedtoinputthecalibrationconstantsforthefibercable.
Thedialogmustbeclosedwithclickingonthe<Finish>button.
6. Thetransmissionofthe
newcalibrationconstantstotheunitisinitializedbyclickingonthe
<DownloadCalibrationConstants>button.Attention:Donotinterruptthedatatransmission!
7. Theclickonthe<Exit>buttoncompletestherecalibrationofthefibercable.
Tighten only after securing connecting sleeve
Note keyed connecting sleeve (can be
inserted only one way - slot must face mode
switches
)
Push connecting sleeve in until it stops,
approx. 15 mm (0.6 in)
Tighten screw (finger tighten only)