9.4ResponseTimeinSetupMode
The analog output responsetime is not guaranteed whilea parameter value is being changed orif
thereisacontinuousstreamofcommandsfromthehost.
ThedigitalresponsetimespecifieshowquicklytheunitcanreportatemperaturechangeviaRS485in
burst mode. (Digital response time is not
defined for polled mode.) The digital response time is
defined as the time that elapses between a change in target temperature and the transmission of a
burststringreportingthenewtemperature.Actualdigitalresponsetimecanvaryfromonereadingto
thenext,sothedigitalresponsetimeis
definedasthe“averagedigitalresponsetime.”
Theaveragedigitalresponsetimedependsonthenumberofcharactersrequestedintheoutputstring
andwiththebaudrate.Itmaybecomputedasthefollowing:
b
n
t
15000
9.9
+=
where:
t=averageresponsetimeinms
n=thenumberofcharactersinthestringincluding<CR>and<LF>
b=thebaudrate
Example:
Withabaudrateof38400,andanoutputstringcontainingtemperatureunits,2colortemperatureand
ambient(20characters),theaveragedigital
responsetimewouldbethefollowing:
mst 7.17
38400
1500020
9.9 =
+=
Notethat the analog output response time is notaffected by baudrate or the number of characters
transmittedintheburststring.