User manual

SensorLocation
14 MarathonSeriesFA/FR
4.5SensorPlacement(1ColorMode)
Optical head placement for onecolor temperature measurements is more critical than twocolor
measurements.Thesensormusthaveaclearviewofthetarget.Therecanbenoobstructionsonthe
lens, window, or in the atmosphere. The distance from the target can be anywhere beyond the
minimumrequirements,as
longasthetargetcompletelyfillsthefieldofview.Thefollowingfigure
illustratesproperplacementwhenusingtheonecolormode.
Figure10:ProperSensorPlacementin1ColorMode
4.6SensorPlacement(2ColorMode)
Thefollowingfigure showsheadplacementunder variousconditions where twocolor temperature
measurements can be taken. Note, however, that if the sensor signal is reduced more than 95%
(includingemissivityandobscurationofthetarget),thesensoraccuracyalsodegrades.
Tar
g
et
g
reater than spot size
Tar
g
et equal to spot size
Tar
g
et smaller than spot size
best good incorrect