User manual
Maintenance
9.3SensingHeadExchange
MIH sensing heads and electronic boxes can not be
interchanged with MID/MIC sensing heads and
electronicboxes!
The head exchange requires to type in the new sensing head
calibrationdataprintedonthecableasfollows:
1. To exchange the sensing head, disconnect the power of the
unit.
2. Connect the
wires for the new sensing head according to the
colordescriptionontheprintedcircuitboard.
3. SwitchthepowerfortheunittoON.
4. Presssimultaneouslythe<Mode/Down/Up>buttons.
5. Fourcharactersappearinthedisplay(formervalues).Typein
the new desig nator (A) using the <Down/Up> buttons.
Press
the<Mode>button.
6. The second block of four characters appears in the display
(former values). Type in the new designator (B) using the
<Down/Up> buttons. Activate your settings by pressing the
<Mode>button.
B
A
Figure45:SensingHeadCalibrationDataprintedontheCable
(e.g.Headwithtwoblocksof4numbers)
For MIH models and specially modified models (like G5 or MTB),
fourblocksoffourcharactersareused.
AlternativelyyoualsocanusetheDataTempMultiDropsoftwarefor
typinginthenew
sensingheadcalibrationdata.
MID 63