User manual

Maintenance
9.2FailSafeOperation
TheFailSafesystemisdesignedtoalerttheoperatorandprovidea
safe outputincase ofanysystemfailure.The sensorisdesignedto
shutdowntheprocessintheeventofasetuperror,sys temerror,ora
failureinthesensorelectronics.
The FailSafe circuit should
never be relied on
exclusively to protect critical processes. Other safety
devicesshouldalsobeusedtosupplementthisfunction!
When an error or failure does occur, the display indicates the
possible failurearea,and theoutput circuitsautomaticallyadjust to
theirpresetlevels,seethefollowingtables.
ErrorCodes
fortheOutputs
Symptom mV 0 to 20 mA 4 to 20 mA TC-K TC-J
Temperature over
range
5 V 21 mA 21 mA - -
Temperature under
range
0 V 2.5 mA 0 mA - -
Defect of the internal
head ambient
temperature probe
5 V 21 mA 21 mA > 1200°C
(2192 °F)
> 1200°C
(2192 °F)
Table7:ErrorCodes(Outputs)
MID 61