8.8ProtectiveWindow
Theprotectivewindowcanbeusedtoprotectthesensingheadfrom
dustandothercontamination.Thisshouldbe appliedespeciallyfor
sensors without a lens. These are all models with an optical
resolutionof2:1.
The protective window is made from nonpoisonous zinc sulfide,
withatransmissionfactorof
0.75±0.05.Ithasanouterdiameterof
17 mm (0.67 in). The protective window can bedirectlyscrewed to
the sensing head. It withstands ambient temperatures up to 180°C
(356°F).
For correcttemperature readings, the transmission ofthe
protectivewindowmustbesetviathecontrolpanelin
the
sensor’s electronic housing, see section 6.2
Setting of
Modes
onpage35!