4.5ElectricalInterference
To minimize electrical or electromagneticinterferenceor“noise”be
awareofthefollowing:
Mounttheunitasfarawayas possible frompotentialsources
of electrical interference such as motorized equipment
producinglargesteploadchanges.
Useshieldedwireforallinputandoutputconnections.
Makesuretheshield
wiresareearthgroundedatonepoint.
Sensor head shield braid should make direct contact around
thecablecircumference.