- Raytek Corp MI Miniature Infrared Sensor Operating Instructions

Maintenance
MI 63
9.3SensingHeadExchange
Sensing heads and electronic boxes can only be
interchangedinaccordancetothefollowing table!
MID02 MIC02 MID10 MIC10 MIH10 MID20 MIC20 MIH20
MID02 x x x x
MIC02 x x x x
MID10 x x x x
MIC10 x x x x
MIH10 x
MID20 x
MIC20 x
MIH20 x
The head exchange requires to type in the new sen sing head
calibrationdataprintedonthecableasfollows:
1. To exchange the sensing head, disconnect the power of the
unit.
2. Connect the wires for the new sensing head according to the
colordescriptionontheprintedcircuitboard.
3.
SwitchthepowerfortheunittoON.
4. Presssimultaneouslythe<Mode/Down/Up>buttons.
5. Fourcharactersappearinthedisplay(formervalues).Typein
the new desig nator (A) using the <Down/Up> buttons. Press
the<Mode>button.
6. The seco nd block of four characters appears in the display
(former values).
Type in the new designator (B) using the