Manual

Figure 4. kV and kVp
Table: Active compensation on kV measurements
Sensor Filtration kV/kVp
R/F low Max. 1 mm Cu or equivalent kV
R/F high Max. 1 mm Cu or equivalent kVp (when signal level is high enough)
Mammography:
Mo/Mo
25 – 35 µm Mo kV (choose “paddle” or “no paddle”)
Mo/Rh
25 – 30 µm Rh kV (use 2mm Al filter)
W/Rh
55 – 60 µm Rh kV (choose “paddle” or “no paddle”)
W/Al
0.65 – 0.75 mm Al kV (choose ”paddle” or ”no paddle”)
W/Al S (scanning)
0.5 mm Al total filtration kV (use 2mm Al filter)
Use a kVp delay on machines with a slow rising output. 150 ms on dental single phase machines and 1000 ms
on a fluoroscopy machine is recommended.
kV is available for R/F and mammography detectors.
NOTE! When selecting the Mo/Mo W/Rh, or W/Al mammography beam quality, you also get at choice
of “paddle” or “no paddle” which only applies if the kV measurement is going to be made with
or without a paddle in the beam and does not aect dose or HVL measurements.
TIME
The RaySafe Xi measures time from the start trig (may be adjusted with the Trig level setting) until the signal
falls below 25 % of the peak (50 or 75 % if adjusted with the Trig level setting). At low dose rates (about 1 %
of the max dose rate for the active sensor), the 25 % end level is changed to a low level (about the lowest
measurable dose rate for the active sensor).
36
RaySafe Xi User Manual – Measurement definitions