Manual

Dose rate Gy, R or Sv per s, m or h
Dose rate, peak (^) Gy, R or Sv per s, m or h
Accumulated dose Gy, R or Sv
While in measure mode, a short press on the detector key stops measurement.
Displayed in MEASURE MODE Displayed in IDLE MODE
instantaneous dose rate average dose rate
peak dose rate (^) peak dose rate (^)
accumulated dose accumulated dose
A long press on the detector key induces a new oset adjustment, and the instrument returns to IDLE MODE.
3. SETUP MENU
Press SELECT to enter the SETUP MENU (from MEASURE MODE) and STEP to step between setup
parameters.
All values are stored in a non-volatile memory and are valid until manually changed.
(Factory settings in bold.)
Dose unit Change the default units and/or use this feature to recalculate a value to a dierent unit.
(Gy, R or Sv per s, m or h, where 1 Gy = 114.1 R.)
27
RaySafe Xi User Manual – Survey Measurement