SURVEY MEASUREMENT
The RaySafe Xi survey detector is designed for measuring leakage or scattered radiation from X-ray equipment
and γ-emitting isotopes.
Available dose quantities (selectable from SETUP MENU) are:
• Air kerma (K
air
), in Gy
• Exposure (X), in R
• Ambient dose equivalent (H*(10)), in Sv
1. SENSOR MENU
R/F For leakage or scatter from R/F equipment.
Mammography For leakage or scatter from mammography equipment.
Nuclear Med. When measuring on γ-emitting isotopes.
After an oset adjustment, the detector waits in IDLE MODE. Start measuring by a short press on the RaySafe
Xi survey detector key.
2. MEASURE MODE
In MEASURE MODE the ticker frequency is proportional to current dose rate. Measured values are continuously
updated on the display (twice per second).
26
RaySafe Xi User Manual – Survey Measurement