MAMMOGRAPHY MEASUREMENT
Place the detector along the X-ray tube axis (with the cable end pointing towards the X-ray machine). For kV
measurements, make sure that the center of the MAM sensor square is placed 6 cm from the examination table
front edge. Dose and HVL can be measured from other positions as well.
The following concerns Unfors Xi Platinum, not RaySafe Xi:
NOTE! When measuring on mammography machines with carbon fibre table, we recommend not to
use the power adapter to charge Xi, or use something that creates at least a 5mm distance
between the table and the detector, such as a plastic plate.
SPECIAL MEASUREMENT CASES
The table below indicates when to use filter for kV measurements.
Beam quality Filter Comment
Mo/Mo No filter
Mo/Rh 2 mm Al
W/Rh No filter
W/Al 20–39 kV No filter
W/Al 40–49kV 2 mm Al Use the R/F high detector
W/AlScanning 2 mm Al
Mo/Rh
When measuring kV with Mo/Rh, use the included 2mm Al filter. Follow the instructions printed on the filter.
14
RaySafe Xi User Manual – Mammography Measurement