Dose unit Change the default units and/or use this feature to recalculate a value to a dierent unit.
(Gy or R per s, m or h, where 1 Gy=114.1 R.)
Backlight Display lighting option. (o, 2, 5, 10 s, on)
Auto Power O Never, or after 5, 20 or 60 minutes of inactivity.
Info Displays the serial number (S/N) and firmware version of the connected detector and
the RaySafe Xi base unit and the calibration date for the selected sensor. There is also a
possibility to reset the instrument settings to factory defaults.
kVp Mode “kV only” prevents RaySafe Xi from calculating kVp, which otherwise occurs
automatically when R/F high is selected and the signal level is high enough (“kV/kVp”).
Useful for example when measuring near that signal level.
13