The RaySafe Solo detector end trigs during exposure.
Make sure to select an appropriate Calc. delay if using the instrument with pulsed generators, or with
generators having a pre-pulse before the main exposure.
The RaySafe Solo does not respond when pressing keys.
Try the reset switch and make sure that the battery is charged.
I get strange readings when measuring HVL on mammography, in comparison with my manual HVL
measurements.
Please make sure that the thickness of your Al filters is accurately known. A resolution on the order of 1 µm
is needed for 0.1 mm thickness.
I get strange mA/mAs readings when I’m using the RaySafe Solo’s AC adaptor or when my computer is
connected to mains power.
Make sure that the instrument
• was connected to the generator during zero adjust.
• is operating on battery (no AC adaptor).
• is not connected to a grounded computer (use Bluetooth or a laptop running on batteries with no Ethernet
cable connected).