Owner's manual

Figure 5. 75 % trig level activated (figure also applies to 25 and 50 % trig level settings). Please note that
the trig level setting aects the dose measurement.
Single phase generators (normally dental applications) may have slowly increasing amplitude characteristics for
approximately 100 ms. It is then recommended to activate a higher trig level in the SETUP MENU after the first
exposure.
PULSE
The RaySafe Solo counts pulses from 1 to 9999. The trigger to increment the pulse counter occurs when the
dose rate has a negative slope and the amplitude falls below 25 % of the peak amplitude for the exposure. At
peak dose rates of approximately < 3 µGy/s (for R/F low) the signal to noise ratio is too low to count pulses and
“---” will be displayed. If possible, increase the mA and/or kVp to increase the radiation output or decrease the
FDD.
Pulse is available for the R/F, RAD, Dose, DENT and mA/mAs detectors.
HVL HALF VALUE LAYER  OPTIONAL
The RaySafe Solo calculates HVL as a function of the signals from several sensor elements from the whole
exposure, if exposure time is less than 6 seconds.
If the exposure is longer than 6 seconds, a HVL value recorded approximately 2 seconds before end trig will be
displayed.
30
RaySafe Solo User Manual – Measurement definitions