Owner's manual

Info Displays the serial number (S/N) and firmware version of the detector and the RaySafe
Solo base unit and the calibration date for the selected sensor. There is also a
possibility to reset the instrument settings to factory defaults.
NOTE! It is recommended to use a laptop operating on battery or a Bluetooth connection when
capturing waveforms or transferring data to a computer running RaySafe Xi View. Connection
of a PC (mains supplied) via a serial cable to the RaySafe Solo base unit may create ground
currents aecting the measurement.
Be aware of rush currents (loading the capacitance in the high voltage cables) in the initial part of the exposure.
Use a Trig delay of 5 ms in the SETUP MENU to exclude the rush current from the measurement, if desired.
24
RaySafe Solo User Manual – mA/mAs Measurement