Owner's manual

R/F MEASUREMENT
Using R/F, RAD, Dose or DENT version.
For best accuracy, center the selected sensor field (R/F low, R/F high, RAD, DENT low or DENT high) and
position the long axis of the sensor field perpendicular to the anode-cathode axis of the tube.
1. SENSOR MENU
R/F low
(R/F and Dose)
Sensor for conventional low dose rate measurements < 1 mGy/s
(7 R/min), normally after a phantom.
R/F high
(R/F and Dose)
Sensor for conventional high dose rate measurements > 1 mGy/s
(7 R/min), normally before a phantom.
RAD Sensor for conventional high dose rate measurements > 0.1 mGy/s
(0.7 R/min), normally before a phantom.
DENT low Sensor for low dose rate measurements < 1 mGy/s
(7 R/min), normally after a phantom or for applications with high
(> 10 mm Al or equivalent) inherent filtration.
DENT high Sensor for high dose rate measurements > 1 mGy/s
(7 R/min), normally before a phantom.
Press STEP to step between sensors and SELECT to select a sensor.
10
RaySafe Solo User Manual – R/F Measurement