User Manual RaySafe i2 System
© 2013.01 Unfors RaySafe 5001047-C All rights are reserved. Reproduction or transmission in whole or in part, in any form or by any means, electronic, mechanical or otherwise, is prohibited without the prior written consent of the copyright owner.
RaySafe i2 System User Manual – Contents Contents Introduction..................................................................................................... 5 About the RaySafe i2 System.......................................................................... 5 How the system works.................................................................................... 6 About this user manual.................................................................................... 6 Using the Dosimeter......
RaySafe i2 System User Manual – Contents Technical data.................................................................................................. 31 Dosimeter radiology ...................................................................................... 31 Dosimeter memory ........................................................................................ 31 Real time display memory ............................................................................
RaySafe i2 System User Manual – Introduction Introduction About the RaySafe i2 System Figure 1.
RaySafe i2 System User Manual – Introduction How the system works The dosimeter measures and records X-ray exposure every second and transfers the data wirelessly, via radio, to the real time display. The real time display shows real time dose exposure from up to eight dosimeters in range at a time. Color indication bars (green, yellow, red) represents the intensity of the currently received exposure. The accumulated dose per individual is displayed next to the color indication bars.
RaySafe i2 System User Manual – Using the Dosimeter Using the Dosimeter Figure 2. The dosimeter Introduction The dosimeter is an active dose meter designed for maintenance-free usage throughout its lifetime. You can personalize the dosimeter’s appearance by attaching one of the 8 inlays of different color, which are delivered together with the dosimeter. (Select which color to show on the real time display, using the dosimeter options dialog in dose viewer or dose manager.
RaySafe i2 System User Manual – Using the cradle Using the cradle Figure 3. The cradle Introduction The cradle is a dock station that lets you connect a dosimeter to a computer for data read out as well as dosimeter Options writing into the dosimeter. Use the cradle in combination with the dose viewer and/or dose manager applications. Getting started NOTE Do not connect the cradle to the computer unless dose viewer and the cradle driver are installed on the computer. 1.
RaySafe i2 System User Manual – Using dose viewer and dose manager Using dose viewer and dose manager See dose viewer and dose manager user manuals, found in the help menu in the dose viewer and dose manager software.
RaySafe i2 System User Manual – Using the real time display Using the real time display Introduction The real time display is the display unit in the RaySafe i2 system. It communicates wireless with dosimeters within radio range in order to collect, present and store dosimeter dose data. Figure 4. The real time display. WARNING The distance from which the real time display detects a dosimeter depends on the settings that the service engineer made during the installation.
RaySafe i2 System User Manual – Using the real time display The real time display screen is a touch screen. Operate the real time display by tapping buttons and graphs on the screen. NOTE Dosimeters are shown on a first come first serve principle (see “Reserved slots menu” in section “Admin Settings menu” on page 20). dosimeters that are not shown in the Online View still measure and store dose data. WARNING Do not cover the real time display to prevent excessive temperatures.
RaySafe i2 System User Manual – Using the real time display Online view – view current dose rates Figure 5. The online view NOTE Dosimeters are shown on a first come first serve principle (see “reserved slots menu” in section “Admin Settings menu” on page 20). Dosimeters that are not shown in the online view still measure and store dose data (see “Dosimeter memory” on page 31).
RaySafe i2 System User Manual – Using the real time display 1 8 2 3 5 4 7 6 Figure 6. The online view elements Legend 1 Personal dose overview (see page 14) access (tap anywhere on the white bar) 2 Scale, green zone 3 Scale, yellow zone 4 Scale, red zone 5 Accumulated procedure dose 6 Peak dose rate indication 7 Displayed name 8 Displayed symbol (color of the dosimeter) The displayed name identifies the dosimeters. It may be truncated due to lack of space.
RaySafe i2 System User Manual – Using the real time display When a dosimeter moves out of range of the real time display, its button will remain visible in the online view for three minutes. This feature allows people to temporarily leave the room without loosing their position in the list. If no dosimeter has been within range of the real time display for 30 minutes, the online view will enter screen saver mode. In screen saver mode the real time display displays a black screen.
RaySafe i2 System User Manual – Using the real time display • Dosimeter – a unique ID per dosimeter • Battery status • • 2-4 blocks indicate normal state. • 1 block indicates that the dosimeter needs to be replaced. The dosimeter’s date and time The Personal dose overview has two sub views: • the annual dose view (see section “Annual dose” on page 16) • the personal dose history view (see section “Dose history” on page 17) 1 2 2 3 Figure 8.
RaySafe i2 System User Manual – Using the real time display • Dose session 1 and 2 values, date and time. Use the dose sessions to measure dose for specific time spans, for example a specific procedure or a working day. The value shows the accumulated dose for a session since last dose session reset. Use the reset buttons (2) to reset a dose session to zero. Information about dose session reset will appear as an event in the dose manager software. • Total personal dose since the last dose history reset.
RaySafe i2 System User Manual – Using the real time display • The annual dose limit distributed over the full year (dashed line). As long as the accumulated dose stays below this value, the annual dose limit will not be exceeded for the full year. • Accumulated dose in relation to the annual dose limit for the current year (green/orange line). When the accumulated dose exceeds the distributed annual dose limit, the color will change from green to orange in order to alert the user to take actions.
RaySafe i2 System User Manual – Using the real time display In the day/week time spans you can view accumulated dose value bars. Each bar represents the accumulated dose during one/four hours, respectively. Within the day and week time spans, tapping a dose bar in the diagram will zoom into the larger underlying time scale (that is week -> day and day -> hour). Within each time span (hour/day/week) you can step forward and backward in time with the arrow buttons on the bottom of the screen.
RaySafe i2 System User Manual – Using the real time display User Settings menu Figure 11. The user settings menu The user settings menu consists of one screen. Adjust the backlight level of the screen by moving the arrow horizontally. This change will take effect immediately.
RaySafe i2 System User Manual – Using the real time display Admin Settings menu The Admin Settings menu consists of four screens. Access these screens by tapping the arrow buttons in the lower right corner of each screen. Real time display name and location menu Figure 12. The real time display name and location menu Enter name and location of the real time display. This information is used to identify the real time display so that it can be detected in dose manager.
RaySafe i2 System User Manual – Using the real time display Reserved Slots menu Figure 13. The reserved slots menu This is an optional setting, which you can use to reserve a number of slots in the online view for the clinical roles doctor, nurse and technician (see figure above). If you, for example, have reserved three slots for doctors but the real time display detects only one, there will be two empty slots before the first nurse appears.
RaySafe i2 System User Manual – Using the real time display Network Setup menu Figure 14. The network setup menu Configure the network connection between the real time display and the dose manager (see figure above). You might have to contact the local IT department to receive the IP address. If you change this information, the real time display requires a restart. NOTE Network setup is only applicable if you are using the dose manager software.
RaySafe i2 System User Manual – Using the real time display Time and Date menu Figure 15. The time and date menu Set time, date, time zone and manage daylight saving time for the real time display (see figure above). For daylight saving time, you can select to manage it manually or by selecting a daylight settings region for your time zone. The possible regions available for daylight saving time are different depending on your current time zone.
RaySafe i2 System User Manual – Maintenance Maintenance Calibrating the real time display screen If the touch screen does not respond correctly to user interaction, a touch screen re-calibration may be needed. Follow the instructions below to perform a re-calibration: 1. Tap and hold on the start up screen that appears during real time display startup until the setup view appears. 2. Follow the instruction on the bottom of the screen and the subsequent instructions in the next views.
RaySafe i2 System User Manual – Maintenance Chrome parts should only be cleaned by rubbing down with a dry cloth. Do not use abrasive polishes. To preserve the finish, use non-abrasive wax. Disinfection Those parts of the product that are suitable for such treatment, including accessories and connecting cables, can be disinfected by wiping with a cloth dampened with a suitable agent. Never use corrosive or solvent disinfectants. If you are not sure about the properties of a disinfectant, do not use it.
RaySafe i2 System User Manual – Troubleshooting Troubleshooting Real time display Table 1. Real time display troubleshooting Problem Solution The dosimeter does not appear in the online view. Use dose viewer or dose manager to check that the dosimeter’s: • Battery status is OK. • Power mode is set to ‘On’. • Real time display mode is set to ‘Show’. Read how to change dosimeter settings in the dose viewer or dose manager user manual. Dose viewer, dose manager and cradle Table 2.
RaySafe i2 System User Manual – Troubleshooting Problem Solution Dose viewer/dose manager does not detect the cradle • Check the USB connection between the cradle and the computer. • Install the cradle driver manually. All the driver files are located on the installation CD in the folder “CradleDriver”. These files are also copied to the application installation folder when dose viewer is installed.
RaySafe i2 System User Manual – Important information About the system Important information About the system Intended use of the RaySafe i2 System This product is intended to be installed, used and operated only in accordance with the safety procedures and operating instructions given in these Instructions for Use for the purpose for which it was designed. The purpose for which the product is intended is given below.
RaySafe i2 System User Manual – Important information About the system or operation, may relieve the manufacturer (or his agent) from all or some responsibility for resultant noncompliance, damage or injury. Compatibility The product described in this manual should not be used in combination with other products or components unless such other products or components are expressly recognized as compatible by Unfors RaySafe. (A list of such products and components is available from the manufacturer.
RaySafe i2 System User Manual – Important information About the system System requirements Dose viewer and dose manager has the following system requirements: • Operating systems: Windows 7, Windows Vista or Windows XP • .NET 3.0 (dose manager: .NET 3.
RaySafe i2 System User Manual – Technical data Technical data Dosimeter radiology Characteristic Measure Operational Quantity* Hp(10) Reproducibility 10% or 1 µSv, whatever is greatest Dose Rate Range 40 µSv/h – 300 mSv/h Energy dependence X-/γ-rays ± 20% within N40 – N100 ± 30% within N100 – N120 Angular dependence ± 5% within ± 5° ± 30% within ± 50° +200%/-100% within ± 90° Temperature dependence ± 5% within 20-26°C ± 25% within 15-35°C Battery voltage dependence ± 2% from fully charged un
RaySafe i2 System User Manual – Technical data • The dose rate memory, where the dosimeter stores dose rate samples. When the radiation exceeds 40 µSv/h, the dosimeter stores one sample per second. This memory is limited to 3600 second-samples. NOTE When the dose rate memory is full, the oldest data will be overwritten with newer data. Real time display memory The real time display stores dose information and dosimeter information for dosimeters that have been connected to it.
RaySafe i2 System User Manual – Technical data Figure 16. Dose viewer/dose manager chart. In lack of dose rate samples, the yellow curve displays mean dose rate values per hour calculated from the accumulated dose (blue curve). NOTE The same effect as illustrated in the figure above is also obtained by un-checking the show details check box in the data filter panel. Time management The dosimeter logs dose history in local time with no daylight saving time adjustments.
RaySafe i2 System User Manual – Technical data Real time display Table 3. Real time display technical specifications Characteristic Measure Weight 1.2 kg (2.7 lbs) Dimensions 297x243x51 mm (WxHxD) Display 10.
RaySafe i2 System User Manual – Technical data Characteristic Measure Log memory 2 3600 Dose Rate samples with one-second resolution, cyclically overridden.
RaySafe i2 System User Manual – Technical data Climatic conditions Classes: C1 Indoors, temperature controlled Parameter Measure Temperature levels: – for performance +15°C to +35°C – for safety +10°C to +40°C – for storage -25°C to +70°C Temperature rate of change < 0.
RaySafe i2 System User Manual – Technical data Security and privacy requirements Anti virus An AVS solution is not installed on this Windows CE device because of the security measures taken to reduce the attack surface. Network ports The following ports and protocols are open on the real time display for communication with the dose manager: TCP/UDP Port number Protocol Additional note TCP 8070 gSOAP httpd 2.
RaySafe i2 System User Manual – Safety Safety Important safety directions If the RaySafe i2 system is not functioning correct or damage is visible, please contact your Unfors RaySafe representative, who will take appropriate actions in order not to harm personnel or patients. Handle the hardware and software with care. Make sure that the hardware and software is used and stored in a secured environment to prevent unauthorized access.
RaySafe i2 System User Manual – Safety Safety devices Warning Never attempt to remove, modify, override or frustrate any safety device on the product. Interfering with safety devices could lead to fatal or other serious personal injury. Intended use and compatibility Warning Do not use the product for any purpose other than those for which it is intended. Do not use the product with products other than that which Unfors RaySafe recognizes as compatible.
RaySafe i2 System User Manual – Safety Fire safety Use of electrical product in an environment for which it was not designed can lead to fire or explosion. Fire regulations for the type of medical area being used should be fully applied, observed and enforced. Fire extinguishers should be available for both electrical and non-electrical fires. WARNING Only use extinguishers on electrical or chemical fires, which are specifically labeled for those purposes.
RaySafe i2 System User Manual – Safety Electromagnetic Compatibility (EMC) This Unfors RaySafe product complies with relevant international and national laws and standards on EMC (electromagnetic compatibility) for this type of product when used as intended. Such laws and standards define both the permissible electromagnetic emission levels from equipment and its required immunity to electromagnetic interference from external sources.
RaySafe i2 System User Manual – Safety against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
RaySafe i2 System User Manual – Safety Network safety, security and privacy Customer Role in the Product Security Partnership We recognize that the security of Unfors RaySafe products is an important part of your facility’s security-in depth strategy. However, these benefits can only be realized if you implement a comprehensive, multi-layered strategy (including policies, processes, and technologies) to protect information and systems from external and internal threats.
RaySafe i2 System User Manual – Safety Explanation of the label symbols Symbol Explanation Caution, consult accompanying documents Conformity marking to applicable European directives IP21 Ingress protection code for the product housing Electrical safety compliance marking, together with listing number Conformity marking for Australian electromagnetic compatibility and radio requirements FCC Radio certification number for US market IC Radio certification number for Canadian market R Radio certifi
RaySafe i2 System User Manual – Safety Real time display product label The following real time display product label is located on the rear side of the real time display. Raysafe i2 real time display D a te o f M fg: Model: 8603025 0 1 /2 0 1 0 1.2 A INPUT RATING: 12 V 918.3 M H z S/N: 0000 0000 IP21 T his devices com plies w ith part 15 of the F C C rules.
RaySafe i2 System User Manual – Safety Raysafe i2 real time display Model: 8603024 INPUT RATING: 8 6 8 ,3 M H z S/N: 0000 0000 D a te o f M fg: 0 1 /2 0 1 0 12 V IP21 1.2 A Designed by Unfors Raysafe Made in PRC Figure 18. Real time display label for products sold in the EU Raysafe i2 real time display Model: 8603027 INPUT RATING: 927,9 M H z D a te o f M fg: 0 1 /2 0 1 0 12 V S/N: 0000 0000 1.2 A IP21 R 008120016 Designed by Unfors Raysafe Made in PRC Figure 19.
RaySafe i2 System User Manual – Safety Date of Mfg: 01/2010 918,3 MHz FCC ID: XWK8601015 Electrical Safety E113017 IC: 9038A-8601015 Model: 8601019 S/N: 0000 0000 RaySafe i2 dosimeter Designed by Unfors RaySafe Made in PRC Figure 21. Dosimeter label The following dosimeter product label is located on the rear side of the dosimeter for products sold in European Union (EU) countries.
RaySafe i2 System User Manual – Product disposal Product disposal Introduction Unfors RaySafe is concerned to help protect the natural environment, and to help ensure continued safe and effective use of this product, through proper support, maintenance and training. Therefore Unfors RaySafe products are designed and manufactured to comply with relevant guidelines for environmental protection. As long as the product is properly operated and maintained, it presents no environmental risks.
RaySafe i2 System User Manual – Product disposal In the European Union (the WEEE directive), this label indicates that this product should not be disposed of with household waste. This product should be disposed of at an appropriate facility to enable recovery and recycling. Unfors RaySafe supports users in: • Recovering reusable parts. • Recycling of useful materials by competent disposal companies. • Safe and effective disposal of product.