Specifications

25
Chapter 4: Calibration
4.1 Introduction
Before using the Rudder Angle Indicator instrument for navigational purposes, it
must be aligned with the Rudder Reference Transducer to counter any offset and
ensure that the rudder position is accurately monitored.
4.2 Rudder angle offset adjustment
Rudder angle offset adjustment removes any deviation between the reading on
the ST60+ Rudder Angle Indicator display and the Rudder Reference Transducer.
To adjust for any offset, proceed as follows:
1. Center the rudder and check that the display analog pointer is within ±7° of
zero. If the analog pointer reading exceeds the limits ±7°, the Rudder Refer-
ence Transducer should be realigned in order to avoid scale inaccuracies (see
Chapter 3, Installation
).
2. With the rudder centered, press and hold the
set button for 2 seconds. The
instrument will emit an audible beep to indicate that you are in calibration
mode.
Note:
You can enter calibration mode only when the instrument is in master mode. If the
instrument is in repeater mode and you attempt to enter calibration mode, the instrument
will beep twice when the
set
button is released, to indicate an invalid button press.
3. Use the < (decrement) and/or > (increment) buttons to adjust the Rudder
Angle Indicator display pointer to the zero position in 1° steps, or press and
hold either
< or > for rapid change.
4. Press the
set button for 2 seconds to save the new setting and exit calibration
mode.
Note:
If, during calibration, no button is pressed for a period of 60 seconds, calibration
mode is exited without change.
81266_1.book Page 25 Monday, December 5, 2005 9:28 AM