5. Maintenance
5.1 Cleaning
1. Clean the detector with IPA (Isopropyl-alcohol) when it is contaminated.
2. Before cleaning the detector, turn off the power and separate the battery.
3. Wear waterproof gloves to protect your hands from direct contact with IPA or any other
liquid.
4. Do not pour or spray IPA directly on the detector. Use fabric or soft cloth moistened with IPA
to clean.
5. Avoid getting IPA or any other liquid into the detector.
6. After cleaning, wait until the IPA is dried completely.
5.2 Inspection
1. In order to ensure that the detector is used safely and normally, please be sure to inspect
the product regularly before use. If any problem occurs, please contact Rayence Customer
Service team.
2. Please perform inspections based on the check list below.
Inspection List
U
s
e
r
Vendor Cycle
Check if cables are not
damaged
o
Daily
Check if plugs and connectors
are not loose or damaged
o
Daily
Check if cover or part is not
damaged
o
Daily
Check the LED indicator o
Daily
Re-Calibration o Half Year
Check the performance of the
product by doing test shots
with Phantom or resolution
chart
o
Yearly