1417WGC/WCC 91
3. Open “_vadav.lnk” from “C:\davinci”.
Once the program is opened and the detector is connected, the LINK LED light will blink
and the Detector Status will display detector information as below.
Once the correct Sensor IP is put into the Davinci, it will automatically pull the
parameter of the connected detector.
Model type Parameter Selected
1417WCC [47]1417WCC-R.par
1417WCC_A [69]1417WCC-R_A.par
1417WCC_140 [71]1417WCC-R_140.par
1417WGC [48]1417WGC-R.par
1417WGC_A [70]1417WGC-R_A.par
1417WGC_140 [72]1417WGC-R_140.par
Default IP address for wireless connection is 2.2.2.100 and for wired connection is
2.2.2.101. If the IP address needs to be changed, please refer to 2.1 Detector IP
Address Set Up in Part.2 Service Manual.
In the AP mode, select "AP" from "Wireless Link Quality" and type the detector's
serial number at the "SSID".
If the detector does not communicate with the PC, please check the connection of the cable,