User's Manual Part 1

PART I. User & Installation Manual
56 1417WGC/WCC
Set up SW
1. Connect Detector and turn on the power.
2. Choose Station Mode by pressing and holding the Mode Button.
LED LED Color Mode
MODE
Orange Station Mode(Wireless)
Green AP Mode(Wireless)
None Wired Mode
Blinking Orange and Green
alternatively
Sleep Mode
3. Open “_vadav.lnk” from “C:\davinci”.
Once the program is opened and the detector is connected, the LINK LED light from the
detector will blink and the Detector Status will display information of the detector as below.
Once the correct Sensor IP is put into the Davinci, it will automatically pull the
parameter of the connected detector.
Model type Parameter Selected
1417WCC [47]1417WCC-R.par
1417WCC_A [69]1417WCC-R_A.par
1417WCC_140 [71]1417WCC-R_140.par
1417WGC [48]1417WGC-R.par
1417WGC_A [70]1417WGC-R_A.par
1417WGC_140 [72]1417WGC-R_140.par