User's Manual
Table Of Contents
- Preface
- Contents
- 1. Safety Information
- 2. Product Introduction and Specification
- 2.1 Product Features
- 2.2 Product Components
- 2.3 Part Names and Functions
- 2.4 Part Specifications
- 2.4.1 Detector
- 2.4.2 Battery [Model name: RB37WHA]
- 2.4.3 Mobile charger [Model name: RMC001A]
- 2.4.4 Mobile charger Adaptor [Model name: AFM60US18]
- 2.4.5 Interface box (Optional) [Model name: RAP001A]
- 2.4.6 Power Supply (Optional) [Model name -RP003A]
- 2.4.7 AGI Box (Optional)
- 2.4.8 IrDA module (Optional) [Model name – RI001A]
- 2.4.9 Battery Charger (Optional) [Model name: RC120WA]
- 2.4.10 Battery Charger Adapter (Optional) [Model name: PMP120-13-3]
- 2.4.11 Cable
- 2.5 Environmental Requirements
- 2.6 Dimensions (Unit: mm)
- 3. Installation and Calibration
- 4. Usage
- 5. Maintenance
- 6. Warranty
PART II. Service Manual
1417WGC/WCC 91
3.2 Calibration
In order to properly acquire images, calibration must be performed. Without calibration, optimum
images cannot be acquired.
Rayence recommends to warm up the detector for 5 minutes after turning on the
power.
Calibration should be performed at regular intervals, typically once every six (6)
months, or whenever the central beam of the X-ray source has been moved relative to
the Detector.
3.2.1 Auto Calibration Mode
1. Connect the detector and turn the power on.
2. Open “_vadav.lnk” from “C:\davinci”.
3. Once the detector is connected, information of the detector is displayed in Detector Status
and Link & Ready are checked as below.
If "Detector Status" does not show anything, please refer to 3.1Installation in
Part.1 User & Installation Manual to connect the detector properly.