User's Manual
Table Of Contents
- Preface
- Contents
- 1. Safety Information
- 2. Product Introduction and Specification
- 2.1 Product Features
- 2.2 Product Components
- 2.3 Part Names and Functions
- 2.4 Part Specifications
- 2.4.1 Detector
- 2.4.2 Battery [Model name: RB37WHA]
- 2.4.3 Mobile charger [Model name: RMC001A]
- 2.4.4 Mobile charger Adaptor [Model name: AFM60US18]
- 2.4.5 Interface box (Optional) [Model name: RAP001A]
- 2.4.6 Power Supply (Optional) [Model name -RP003A]
- 2.4.7 AGI Box (Optional)
- 2.4.8 IrDA module (Optional) [Model name – RI001A]
- 2.4.9 Battery Charger (Optional) [Model name: RC120WA]
- 2.4.10 Battery Charger Adapter (Optional) [Model name: PMP120-13-3]
- 2.4.11 Cable
- 2.5 Environmental Requirements
- 2.6 Dimensions (Unit: mm)
- 3. Installation and Calibration
- 4. Usage
- 5. Maintenance
- 6. Warranty
PART II. Service Manual
1417WGC/WCC 81
3. Open “_vadav.lnk” from “C:\davinci”.
Once the program is opened and the detector is connected, the LINK LED light will blink
and the Detector Status will display detector information as below.
Once the correct Sensor IP is put into the Davinci, it will automatically pull the
parameter of the connected detector.
Model type Parameter Selected
1417WCC _127A [69]1417WCC_127A.par
1417WCC _140A [71]1417WCC_140A.par
1417WGC _127A [70]1417WGC_127A.par
1417WGC _140A [72]1417WGC_140A.par
Default IP address for wireless connection is 2.2.2.100 and for wired connection is
2.2.2.101. If the IP address needs to be changed, please refer to 2.1 Detector IP
Address Set Up in Part.2 Service Manual.
If the detector does not communicate with the PC, please check the connection of the
cable, PC set up and power of the detector.