User's Manual

Table Of Contents
PART II. Service Manual
1417WGC/WCC 68
3. Open “_vadav.lnk” from “C:\davinci”.
Once the program is opened and the detector is connected, the LINK LED light will blink
and the Detector Status will display detector information as below.
Once the correct Sensor IP is put into the Davinci, it will automatically pull the
parameter of the connected detector.
Model type Parameter Selected
1417WCC _127A [69]1417WCC_127A.par
1417WCC _140A [71]1417WCC_140A.par
1417WGC _127A [70]1417WGC_127A.par
1417WGC _140A [72]1417WGC_140A.par
Default IP address for wireless connection is 2.2.2.100 and for wired connection
is 2.2.2.101. If the IP address needs to be changed, please refer to 2.1 Detector
IP Address Set Up in Part.2 Service Manual.
If the detector does not communicate with the PC, please check the connection of the
cable, PC set up and power of the detector.