User's Manual

Table Of Contents
PART II. Service Manual
1417WGC/WCC 55
If the detector does not communicate with the PC, please check the connection of the
cable, PC set up and power of the detector.
4. After checking connectivity, click the “Calibration & Acquisition” tab and click "Edit".
5. Another window will now be opened as shown below. Select "Auto Trigger" from "Trigger
Mode". If the "Window time" needs to be changed, type the value at "Window Time" from
"Setting".
In Auto trigger mode, be sure to set the "Window time" longer than an exposure
time. If the "Window time" is shorter than the exposure time, images will not be
properly acquired.