User's Manual
Table Of Contents
- Preface
- Contents
- 1. Safety Information
- 2. Product Introduction and Specification
- 2.1 Product Features
- 2.2 Product Components
- 2.3 Part Names and Functions
- 2.4 Part Specifications
- 2.4.1 Detector
- 2.4.2 Battery [Model name: RB37WHA]
- 2.4.3 Mobile charger [Model name: RMC001A]
- 2.4.4 Mobile charger Adaptor [Model name: AFM60US18]
- 2.4.5 Interface box (Optional) [Model name: RAP001A]
- 2.4.6 Power Supply (Optional) [Model name -RP003A]
- 2.4.7 AGI Box (Optional)
- 2.4.8 IrDA module (Optional) [Model name – RI001A]
- 2.4.9 Battery Charger (Optional) [Model name: RC120WA]
- 2.4.10 Battery Charger Adapter (Optional) [Model name: PMP120-13-3]
- 2.4.11 Cable
- 2.5 Environmental Requirements
- 2.6 Dimensions (Unit: mm)
- 3. Installation and Calibration
- 4. Usage
- 5. Maintenance
- 6. Warranty
PART II. Service Manual
1417WGC/WCC 45
3.1.2 Install battery
Attach the battery (RB37WHA) to the detector as below.
• Battery removal
Push the battery unlock button.
3.1.3 Mode Selection
Mode Description
<Auto Trigger>
Automatically detects X-ray radiation without integration of the
generator and detector.
<Manual Trigger>
Detects X-ray radiation by sending and receiving sync signals through
the integration between generator and detector.
<Station Mode>
Communicates with the wireless AP, and the wireless AP
communicates with the PC through the LAN cable.
(Rayence does not provide wireless AP.)
<AP Mode>
Detector communicates with the PC without the wireless AP.
(PC must maintain wireless network card.)
<Wired Mode>
Communicates with the power supply, and the power supply
communicates with the PC through the LAN cable.