User's Manual
Table Of Contents
- Preface
- Contents
- 1. Safety Information
- 2. Product Introduction and Specification
- 2.1 Product Features
- 2.2 Product Components
- 2.3 Part Names and Functions
- 2.4 Part Specifications
- 2.4.1 Detector
- 2.4.2 Battery [Model name: RB37WHA]
- 2.4.3 Mobile charger [Model name: RMC001A]
- 2.4.4 Mobile charger Adaptor [Model name: AFM60US18]
- 2.4.5 Interface box (Optional) [Model name: RAP001A]
- 2.4.6 Power Supply (Optional) [Model name -RP003A]
- 2.4.7 AGI Box (Optional)
- 2.4.8 IrDA module (Optional) [Model name – RI001A]
- 2.4.9 Battery Charger (Optional) [Model name: RC120WA]
- 2.4.10 Battery Charger Adapter (Optional) [Model name: PMP120-13-3]
- 2.4.11 Cable
- 2.5 Environmental Requirements
- 2.6 Dimensions (Unit: mm)
- 3. Installation and Calibration
- 4. Usage
- 5. Maintenance
- 6. Warranty
PART II. Service Manual
1417WGC/WCC 114
5. Maintenance
5.1 Cleaning
1. Clean the detector with IPA (Isopropyl-alcohol) when it is contaminated.
2. Before cleaning the detector, turn off the power and separate the battery.
3. Wear waterproof gloves to protect your hands from direct contact with IPA or any other
liquid.
4. Do not pour or spray IPA directly on the detector. Use fabric or soft cloth moistened with IPA
to clean.
5. Avoid getting IPA or any other liquid into the detector.
6. After cleaning, wait until the IPA is dried completely.
5.2 Inspection
1. In order to ensure that the detector is used safely and normally, please be sure to inspect
the product regularly before use. If any problem occurs, please contact Rayence Customer
Service team.
2. Please perform inspections based on the check list below.
Inspection List User Vendor Cycle
Check if cables are not damaged o Daily
Check if plugs and connectors are not loose or
damaged
o
Daily
Check if cover or part is not damaged o Daily
Check the LED indicator o Daily
Re-Calibration o Half Year
Check the performance of the product by doing test
shots with Phantom or resolution chart
o
Yearly