PART II. Service Manual
1417WGC/WCC 108
4.4.7 Image Storage function
1. Press the "Mode select button" twice (within 2 seconds) and click one more time to go to
"storage mode".
Mode status can be checked by the OLED window of Detector
2. Press the mode select button for 3 seconds to check the status of storaged images.
When Storage mode is used if the connection between the detector and PC,
the image is stored the memory in the Detector internal
When Mobile sevice connects to Detector, can check correctly Image by
Storage Viewer Web Server
For Storage Viewer, must running DBMS and Web Server envirnoment
If disable ‘0x1e’ parameter by VADAV, running DBMS and Web Server
environment durning Detector boot up
If enable ‘0x1e’ parameter by VADAV, running DBMS and Web Server
environment during enter Storage Mode menu
Parameter ‘0x1e’ is Storage Function Disable
For running Storage Viewer environment, OLED is displayed “Check DB
Server”
Stored Image can be opened and deleted by referring SDK