User's Manual
37
R
-
USM
-
020
B. Mode 2 : Trigger cable mode
Connect the Trigger cable between the X-ray enable connector of
AGI with X-ray generator.
Figure 8: <2nd example> wiring for the interface between portable generator & detector
Connection description
Signal New Label Old Label Color Input / Output
READY IN
12V~24VDC R.E.C. +12-24V Red Input
READY IN Ready Input Signal Black Input
READY
DONE
READY DONE Exposure Output SW1 Green Output
READY DONE COM
Exposure Output SW2 Blue Output
Table 3 : Connection description
Make
assurance
doubly sure
SIGNAL RATING
before connection.
AGI