Instruction manual

Outline of Trigger Recording
4-20
Outline of Trigger Recording
In addition to manual operation, you can also start or stop recording automatically in the following modes.
Repeat Mode
Repeats the recording operation for a specified number of times, as shown in the following diagram. You can
also specify “1” to record one time only.
<Recording-Start Conditions>
Level trigger
The trigger is a change in level in a specified channel.
External trigger
The trigger is a contact signal to the EXT TRIGGER IN the connector on the rear of the LX. (L level: starts if
0.6 V or less) If the level is already L when the system enters the recording-standby status, recording starts
at once in memory/MO recording. In PC recording, however, recording starts when the signal changes from H
to L after the system enters the recording-standby status.
Time-out
Recording is forced to start if the specified recording- start condition is not satisfied during a set period.
Pre-trigger
A pre-trigger also records the data that was read into a buffer before one of the above recording-start
conditions was satisfied. The voice memo is not recorded during that period.
<Recording-Stop Conditions>
Level trigger
The trigger is a change in level in a specified channel.
External trigger
The trigger is a contact signal to the EXT TRIGGER IN the connector on the rear of the LX. (H level: stops if
open or 2 V or more)
Recording time
Records for a specified period only.
Post-trigger
Continues to record for a set period even after a recording-stop condition is satisfied. If the recording is
stopped by manual operation, the post-trigger is not recorded.
To cause a trigger action, click
i the toolbar or press the REC button
on the main unit to put the system in the
recording-standby status.
Even if a trigger action is enabled, the
buttons on the toolbar and main unit are
not disabled.
Start
Stop
Pre
-
trigger
P
ost
-
trigger
Recording start condition is
satisfied:
Level trigger
External trigger
Time out
Recording strop
condition is satisfied:
Level trigger
External trigger
Recording time