Instruction manual

Zero Balance
4-16
Zero Balance
Whenever turning on the power, use the zero balance function at the ST amp.
1.
Click on the toolbar at the stop state of the LX main unit.
2.
In Zero Balancing dialogue, turn on the check box of the channel for zero balance and click the
“Execute” button to start zero balance.
3.
While in zero balance, the following dialogue is displayed and then disappears after showing the result.
4.
The result of each channel’s zero balance is displayed on items of the “State”. The “OK” means normal
and the “NG” means abnormal. Establish a setting of range value at the maximum of
100000uST/50mV/V on the channel with no input signal.
5.
After the zero balance is completed, click Exit or click to return to the main window.