Instruction manual
Sensitivity Setting Using TEDS Functions
4-12
Sensitivity Setting Using TEDS Functions
When a transducer/sensor conforms to the IEEE Std 1,451.4 A SMART TRANSDUCER INTERFACE is
connected to the AR-LXPA100. The LX can read its Transducer Electronic Data Sheet (TEDS) information and
then can display the information and set the acquired sensitivity coefficient automatically to the channel to
which the transducer is connected.
■ If the information of the transducer does not comply with the IEEE TEDS standard template(ver0.9), the
LX cannot read and display the information correctly.
1.
Click the toolbar icon in the main window.
2.
The TEDS dialog is displayed.
If you turn ON the power of the LX after connecting the TEDS transducers, the TEDS information is
automatically loaded at the initial calibration stage, and the LX displays the following window shown below.
3.
Click Update to read the TEDS transducer information.
If you turn ON the power of the LX after connecting the TEDS transducers, the TEDS information is
automatically loaded at the initial calibration stage, and the LX displays this window.