Input Amp Settings
4-6
Input Amp Settings
In the System dialog, click the tab of the slot in which the input amp is installed.
The setting parameters are valid, depending on the type of the amps installed.
DC Amp AR-LXDC100
<DC100> is shown at the Slot tab. See Section 6 for the specifications.
Range Settings
You can select the value from ±0.5、1、2、5、10、20、50 V. A range of +/- 131% of the selected range can be
obtained for AD (analog to digital) conversion value. However, the input margin level is approximately +/-
120%. Select a value so that the input does not exceed ±120%. Input signal, and create the setting by
clicking the right
button to check the level.
Auto
When you input the signal and click Auto, the optimum input range is set automatically.
Select input range
Confirm input level
Confirm input level of every channel
Stop confirming inpu
t
level
Set input range automatically