Instruction manual
System Settings
4-5
Monitor/Generator
In Signal, select the monitor channel or the type of the generator output signals (only for the LX-120) to be
output from the MON OUT connector on the rear panel. If you don't want to monitor any channel, set NONE.
Click Property to set the output range of the monitor channel or the details of the generator output signal.
In Signal, select the monitor channel or the type of the generator output signals (sine, sweep sine, pulse,
pink noise, or white noise from the MON OUT connector of the LX-120).)
When you select the generator output signal, you can select the following parameters that appear at the
dialog. (The parameters that can be set appear automatically, depending on the selection of the output
signal.)
Sin/Sweep
Start Freq: Set the start frequency (Hz) in the range of 0 to 43000(Hz) when <sine> or <sweep sine>
is selected at Signal. However, the upper limit of the monitor output frequency is 1/2.4 of the sampling
frequency.
Stop Freq: Set the sweep stop frequency (Hz) in the range of 0 to 43000(Hz) when <sweep sine> is
selected at Signal. However, the upper limit of the monitor output frequency is 1/2.4 of the sampling
frequency.
Rate: Set the sweep rate (Hz) in the range of 0 to 80(Hz) when <sweep sine> is selected at Signal. If
the value is too much, it might take a long time for monitor output of the frequency specified at Start
Freq and Stop Freq.
Pulse
Pulse Width: Set the pulse width in the range of 0 to 65535(msec) when <pulse> is selected at
Signal.
Pulse Interval: Set the pulse interval in the range of 0 to 65535(msec) when <pulse> is selected at
Signal.
You can set output range in Range. You can select the output range of 1 V to 5 V in 0.1 V steps.
See Section 6 for the specifications of the generator output.
■ You cannot use the tachometer pulse input channels and the generator output function at the same time.
■ You cannot use the monitor output of the analog signal when recording at the low-speed sampling series
(1 kHz to 1/60 Hz).
■ The available frequency of sine or sweep sine for generator output is 1/2.4 of the sampling frequency.