Instruction manual

Sampling Frequency and Number of Channels
1-14
(16bitsAD/24bitsAD)
65.536kHz series
Max. number of analog input channels based on the tachometer input setting
Sampling Frequency(kHz)
Bandwidth (kHz)
No 16bits 2ch 32bits 1ch 32bits 2ch
*65.536 27 4/2 2/Unrecordable 2/Unrecordable Unrecordable
*32.768 13 8/4 4/Unrecordable 4/2 4/2
16.384 6 16/8 8/Unrecordable 8/4 8/4
8.192 3 32/16 24/Unrecordable 24/8 24/8
4.096 1.7 32/16 32/Unrecordable 32/16 32/16
2.048 0.8 32/16 32/Unrecordable 32/16 32/16
1.024 0.4 32/16 32/Unrecordable 32/16 32/16
Remarks: Cannot select the moving average other than 1 (one) at the Sampling Frequency settings shown
with *(asterisk).
In 24 bits A/D, 2 ch (16 bits) of tachometer input is not available.
When using LAN as the main unit interface, continuous recording operations might not be possible (it
might be stopped while recording because all of the memory has been used) on setting the maximum
number of recording channels. In such a case, decrease the data transfer rate by decreasing the
sampling frequency by half or by decreasing the number of recording channels by half, and use the LX
main unit.