Instruction manual

Generator Output Specifications
6-6
Generator Output Specifications
You can use the MON OUT connector of the LX-120 for the generator output.
Number of Output Channels 1 channel
Output Level 1 to 5 V at 0.1 V steps (same as the monitor output)
Output Connector MON OUT (Monitor output) BNC
Types of Outputs SIN, SWEEP SIN, Pulse, Pink noise, White noise
Noise Level -70 dB or less (at SIN output)
You cannot use the tachometer pulse input channels and the generator output function at the time.
You cannot use the monitor output of the analog signal when recording at the low-speed sampling series
(1 kHz to 1/60 Hz).
The upper limit of frequency of sine, sweep sine for generator output is 1/2.4 of sampling frequency.