Each video line in four output operating mode consists in:
30 inactive pre-scan, (not connected to pixels),
6 dark references,
4 isolation elements, (inactive, not connected to pixels),
3 non-useful pixels,
3 000 useful pixels of the line.
N = number of pixel periods (T
p
) during readout period (see Figure 5).
Four output operating mode: N 3043.
Two output operating mode: N 6086.
(Φ
LS
can be clocked during the line blancking).
Figure 4. Detailed Timing Diagram For Transfer From Photosite To Register
Figure 5. Detailed Pixel Timing Diagram
T
P
= Pixel period
Rise and fall time:
Φ
R1-2
, Φ
R3-4
:5%ofT
P
(min. 5 ns),
Φ
LS1-2
, Φ
LS3-4
:5%ofT
P
(min. 5 ns),
Φ
L1
, Φ
L2
: 25% of TP (min. 30 ns),
Φ
P1-2
, Φ
P3-4
: 100 ns (min 20 ns).
20 ns
N1
100 ns
2 µs
ΦP1-2, ΦP3-4
ΦLS
ΦR
ΦL1, ΦL2
20 ns
30 ns
Reset
Signal
ΦP1-2, ΦP3-4
ΦL1, ΦL2
ΦLS1-2, ΦLS3-4
ΦR1-2, ΦR3-4
VOS (1-2-3-4)
(CCD output signal)
Floating
diode
(Reference level for
correlated double Sampling)
Tp (200 ns Typ.)
Tp/2 Typ.