Owner's manual

7
5440A–IMAGE–10/05
[Preliminary] ATMOS -2M60/2M30
If shutter time is set to > 0:
Readout is suspended between two consecutive frame readout and shutter output signal is acti-
vated (programmable polarity) for a time configurable within 0 to Tsh range by step of T1.
Figure 4-2. Free-run Mode and Shutter Chronogram
The maximum frame rate of 48 frame/s for the 2M60 camera (or 24 frame/s for the 2M30 cam-
era) is given in this mode with a shutter time set to 0.
See register Aperture Shutter Time @ 246H, Internal Register Mapping on page 16.
4.1.3 Triggered Mode
Principle: An external trigger starts the reset of the sensor, then snap and readout of a frame,
the integration time is defined by readout time (+ programmable shutter time, when used).
The trigger event initiates the following sequence:
Stop of readout frame in progress and reset of readout pointer to line 1
Readout of previous dummy frame and integration start of the frame N. The readout data of
previous frame is not forwarded on Camera Link interface (FVAL inactive)
Shutter output signal is activated during a time programmable within 0 to Tsh range by step
of T1
Table 4-1. Shutter Time Values at Free-run Mode
Label Description 2M60 2M30
Tsh Maximum shutter time (ms) 655 1310
T1 Step duration (µs) 10 20
Line 1
Integration
(Frame N
)
Readout
and Rese
t
Line n
Integration
(Frame N)
Frame N
Programmable
Shutter Time
Shutter
Line n Readout
and Reset
Readout Time
Readout
Frame N - 1
Time
out
Line 1