10. Reassemble the sensor module by placing the sensor electrode panel back in place
and firmly pressing the cap back onto the top.
11. Place the sensor module back into the ToxiRAE Pro PID. Make sure the index
points are aligned (it can only go in one way).
12. Carefully place the cap back onto the ToxiRAE Pro PID and tighten the screw.
Note: Always calibrate the ToxiRAE Pro PID after replacing the sensor module.
17.6. Replacing The Fan
If the fan stops working or is damaged, it can be replaced (part number G02-3006-000).
Before replacing the fan, make sure the instrument is turned off.
1. Remove the top cap as shown in Section 15.2 (page 59).
2. Grip the fan with two fingers and pull up.
Match index
points on sensor
module with
ToxiRAE Pro PID