ToxiRAE Pro PID Users Guide
60
3. Once the top cap is removed, install the new top cap assembly.
4. Place the new rubber insert onto the ToxiRAE Pro PID, being careful to align the
insert correctly and that it is not distorted. Also, make sure that you do not press
on the fan.
5. Press the new top cap onto the ToxiRAE Pro until it is fully seated. Check to see
that the pin on the front side of the cap went into the corresponding groove on the
front side of the instrument housing. Then tighten the Phillips screw.