User Manual

Table Of Contents
ToxiRAE Pro PID Users Guide
56
7. Before installing a new battery, visually inspect the contact pins on the ToxiRAE
Pro PID circuit board to make sure they are not bent. If they are, carefully
straighten them before installing the battery. Install the new battery (part number
G02-3004-000), making sure it is inserted fully. Make sure the battery connectors
on the circuit board go all the way into the connector slots in the battery.
8. Inspect to make sure the ribbon connector is still connected, and plug it back in if
it has come unplugged.
9. Place the cover over the compartment by tilting it up and then fitting it in place.
10. Tighte n all four screws by turning them clockwise with a torque-limiting
screwdriver (preferably an electric one). Tighten them in the numerical order
shown here, to a torque rating of 3.54 to 4.43 lbf-in (4.08 to 5.10 kgf-cm).
Note: Incorrectly tightening the screws may compromise the instruments
IP rating.